● CELL SET mode (Setting cell constant)
D-52/53/54/55
Set the cell constant the first time an electrode is
connected to the main unit of the meter.
1.
To enter the CELL SET mode, press the CAL
key while in the Measurement mode.
2.
Change the digit number using the ENTER key.
3.
Press the ▲ and ▼ keys to set the cell constant
written on the electrode label.
Setting range: 0.700 – 1.300
To change the coefficient, use the following
procedure.
When the SI unit system (m
When the former unit system (cm
Temperature coefficient
The default value of the temperature coefficient is set at
2.00%/ºC.
To change this setting, refer to "3.4.11 Temperature
coefficient setting (D-54 )" page 87.
Unit Setting
The default value of unit is S/m (SI unit system).
To change this setting to the former unit system S/cm,
refer to "3.4.10 Conductivity unit setting (D-54 )" page
86.
2 Taking Measurements
2.8 Conductivity measurement (D-54)
-1
) is set:
-1
) is set: