● Auto Hold Measurement mode
D-52/53/54/55
Auto Hold Measurement mode maintains the display of
the value measured when the meter automatically
judges that the measured value has stabilized. Press the
MEAS key with the initial screeen displayed to make
"HOLD" blink on the display. When the measured value
becomes stable, "HOLD" will stop blinking and remain
displayed, and the measured value will remain
displayed. To clear the hold status or "stabilized" value
(when "HOLD" is blinking), press the MEAS key.
Criteria for judging stability
pH, ORP, ION
measurement
Conductivity
measurement
DO measurement
Temperature
measurement
Blinks
: Within ±1 mV variance in potential
after 10 seconds
: Within ±3-digit variance after 10
seconds
: Within ±3-digit variance after 10
seconds
: Within ±2ºC variance after 10
seconds
2 Taking Measurements
2.3 Measurement modes
Displayed
23