NAVI D-25 Instruction Manual page 160

Ph meter
Table of Contents

Advertisement

D-52/53/54/55
electrode until the internal solution nears the
filler port.
Filling the electrode: assembly procedure
1.
Return the silicon tube to its original position and
seal the internal solution filler port (inner tube).
(Make sure that the filler port is completely
sealed.)
2.
If the bodies (internal and external) or the liquid
junction (tip) are dirty, wash them with pure (de-
ionized) water.
3.
Insert the internal body into the external body.
(Make sure that the O-ring is properly in place.)
4.
Return the spring to the top of the internal body.
5.
Hold the rubber sleeve in place by hand, then
twist the electrode protective cap 90º and fit the
rubber sleeve into the electrode protective cap.
6.
Maneuver the parts so that the "HORIBA" logo
faces the same direction as the reference
solution filler port (outer tube), then put the cap
on the external body.
7.
Twist the rubber sleeve 90º and hold it in place.
8.
Use the syringe to fill the electrode with the
specified reference solution.
6 Maintenance and Troubleshooting
6.2 ION electrode maintenance
149