Advertisement

4 RS-232C communications
4.3 On-line operation commands
● Start conductivity cell constant calibration command for-
mat (D-54 )
110
Prefix for the unit of the calibration value
to be set
0: -; 1: m; 2: μ
The set unit determines which unit is
used, either S/m or S/cm.
Calibration value to set
Command code (Conductivity calibration start)
Header
・ This
command
conductivity measurement or calibration are on HOLD.
・ This command starts the calibration of the conductivity
cell constant.
Setting ranges for calibration values
・ When units are μS/m: 0.0 – 999.9 (only one place
to the right of the decimal)
・ When units are mS/m or S/m: 0.0000 – 199.9
・ After calibration is completed, the cell constant/
coefficient will appear on the screen and be stored in
the memory.