Software Limits; Changing Main Channel Software Limit; Determining How The Unit Responds When Output Is Off (Load Type) - KEPCO BOP 6-125MG/ME Quick Start Manual

Bop 1kw-mg/me series
Table of Contents

Advertisement

age when the unit is in voltage mode). Start
with the most significant digit of the desired
value, then use
For fine adjustment press the ADJUST control
in while rotating the knob to modify the least
significant digit.
• Enter the desired value on the keypad using
the number keys. For example, to program the
BOP to 75.8V, press the following keys in
75.8
order
adjustment use
significant digit. To correct the entry before
activation press the CLEAR key to set the
value to zero and start over. When the desired
value is displayed, press ENTER. This causes
the new value to appear at the output and be
applied to the load if the output is enabled.
6. To program the corresponding Protect channel,
Y
U
press
or
tect channel. Then set the value using either of
the two methods described above. If the Protect
Entry setting is set to Independent, separate
entries for the positive and negative protect chan-
nel are possible. Otherwise the value entered is
applied to both positive and negative protect
channels.
NOTE: The BOP can be configured to show
the protection limits as either a single
value that applies to both protection
channels or show individual settings
for positive and negative protection
limits. See Operator's Manual for
details
SOFTWARE LIMITS. Software
programming of the main channel or the Protect
channel beyond the software limit value. Refer to
Operator's Manual for a full explanation of software
limits.

Changing Main Channel Software limit.

This procedure allows the user to determine the
maximum value of voltage or current that can be
programmed.
%
1. Press
from the power-up screen to enter the
General Setup menu, then highlight Max/Min Set-
tings.
!
2. Press
to enter the Max/Min Settings submenu
(Figure 6). (If a Password is required, see Opera-
tor's Manual for instructions.)
KEPCO, INC.  131-38 SANFORD AVENUE  FLUSHING, NY. 11355 U.S.A. TEL (718) 461-7000  FAX (718) 767-1102
12
T
to highlight the next digit.
then press ENTER. For fine
Y
U
, and
to modify the least
as necessary to highlight the Pro-
limits
http://www.kepcopower.com   email: hq@kepcopower.com
3. Highlight the voltage or current max/min value
and press
absolute values (do not use minus sign for nega-
tive limits). Use number keys to change the set-
ting, then
4. When complete, press
#
to abort, or
saving for power-up) and exit.
5. Upon return to the power-up screen, the main
channel (voltage or current) is compared against
the main channel limits in effect. If the main chan-
nel exceeds the limit, it is set to zero.
6. Highlight the ±CPROTECT or ±VPROTECT max/
min value and press
limits are absolute values (do not use minus sign
for negative limits). Use number keys to change
the setting. Press
7. When complete, press
#
or
to abort, or
out saving for power-up) and exit.
Upon return to the power-up screen, the new protec-
tion limit (voltage or current) is compared against the
protection limits in effect. If the new protection limit
setting is below the existing setting for the protection
limit, the protection channel (voltage or current) is
set to zero.
DETERMINING HOW THE UNIT RESPONDS
WHEN OUTPUT IS OFF (LOAD TYPE)
BOP supports three Load Type selections (see Table
5) which determine how the power supply responds
when the output is off: ACTIVE, RESISTIVE and
BATTERY. The Load Type selection does not affect
the settings of the power supply for ON state; it only
affects the main internal reference level and the pro-
prevent
tection levels during the OFF state.
For inductive loads, and especially
superconducting
loads, the inherent offset of the
BOP in the OFF state may generate
significant current in the circuit. A
properly rated switch in parallel
with a resistor must be connected
between the power supply and the
load. The switch must be open and
the BOP front panel LCD must read
0V, 0A before removing or installing
connections between BOP and
load.
228-1692 REV 8
!
to change it. Software limits are
$
to save.
$
to save for power-up,
%
to apply the changes (without
!
to change it. Software
$
%
to save, or
$
to save for power-up,
%
to apply the changes (with-
WARNING
magnet
to abort.
The
type
090418