12.5 Printing

Problem
Print function not available
Results report not being
printed
No personal data in results
report
No bioimpedance parameters
in results report
No height in results report
70 •
Cause
Software versions incompatible
PC printer not switched on
The PC on which the
seca 115
software is installed has not been
switched on.
No network connection established
between device and
seca 115
software
No connection established between
PC software and PC printer
seca 115
On repeat measurement: no seca
patient file assigned
On initial measurement: seca patient
file not created yet
No bioimpedance measurement
performed
No height entered on device
No height transmitted by the
length measuring device
Administrator: use compatible software
version:
• device: min. software version 1.1
• PC software: min. software version 1.4
Switch on PC printer
PC
Switch on PC
Administrator: establish Ethernet
PC
connection
Administrator: establish connection
between
seca 115
printer
Call up
patient
file (see "Searching for a seca patient file"
on page 37)
Call up
patient
file (see "Creating a new seca patient file"
on page 40)
Call up the
bia
measurement (see "Performing a
bioimpedance measurement" on page 31)
Call up the
weight/height
height (see "Entering height manually" on
page 29)
Measure the height of the patient again,
then press the
seca 360°
length measuring device (see "Transmitting
height via seca 360° wireless network" on
page 29)
Remedy
PC software and PC
tab and assign seca patient
tab and create seca patient
tab and perform the
tab and enter
button on the
send/print