Important EMC notices for use in medical environments
• The LMD-X310MT/X550MT needs special precautions regarding EMC and needs to be installed and put into service
according to the EMC information provided in the instructions for use.
• Portable and mobile RF communications equipment, such as cellular phones, can affect the LMD-X310MT/X550MT.
WARNING
The use of accessories and cables other than those specified, with the exception of replacement parts sold by Sony
Corporation, may result in increased emissions or decreased immunity of the LMD-X310MT/X550MT.
Guidance and manufacturer's declaration – electromagnetic emissions
The LMD-X310MT/X550MT is intended for use in the electromagnetic environment specified below. The
customer or the user of the LMD-X310MT/X550MT should assure that it is used in such an environment.
Emission test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic
emissions
IEC 61000-3-2
Voltage
fluctuations/flicker
emissions
IEC 61000-3-3
WARNING
If the LMD-X310MT/X550MT will be used adjacent to or stacked with other equipment, normal operation of the LMD-
X310MT/X550MT under such configurations should be verified via observation.
4
Compliance
Group 1
Class B
Not applicable (LMD-X310MT)
Class D (LMD-X550MT)
Not applicable (LMD-X310MT)
Complies (LMD-X550MT)
Electromagnetic environment – guidance
The LMD-X310MT/X550MT uses RF energy only for
its internal function. Therefore, its RF emissions are
very low and are not likely to cause any interference
in nearby electronic equipment.
The LMD-X310MT/X550MT is suitable for use in all
establishments, including domestic establishments
and those directly connected to the public low-voltage
power supply network that supplies buildings used for
domestic purposes.