Sony LMD-X2705MD Instructions For Use Manual

Sony LMD-X2705MD Instructions For Use Manual

Hide thumbs Also See for LMD-X2705MD:

Advertisement

Quick Links

4-747-116-14(1)
2019-12
LCD Monitor
Instructions for Use
Before operating the unit, please read this manual thoroughly
and retain it for future reference.
LMD-X2705MD
LMD-X2700MD
© 2019 Sony Corporation

Advertisement

Table of Contents
loading

Summary of Contents for Sony LMD-X2705MD

  • Page 1 4-747-116-14(1) 2019-12 LCD Monitor Instructions for Use Before operating the unit, please read this manual thoroughly and retain it for future reference. LMD-X2705MD LMD-X2700MD © 2019 Sony Corporation...
  • Page 2 Indications for Use/Intended Use The LCD Monitor is intended to provide 4K 2D color Symbols on the product video displays of images from endoscopic/laparoscopic Safety sign camera systems and other compatible medical imaging Follow the warnings in the instructions for use systems.
  • Page 3 – Relocate the unit with respect to the affected devices. – Connect the unit and the affected devices to different branch circuits. For more information, consult qualified Sony service personnel. (Applicable standard: IEC 60601-1-2)
  • Page 4 LMD-X2705MD/X2700MD under such configurations should be verified via observation. • The use of accessories and cables other than those specified, with the exception of replacement parts sold by Sony Corporation, may result in increased emissions or decreased immunity of the LMD-X2705MD/X2700MD.
  • Page 5 Guidance and manufacturer’s declaration – electromagnetic immunity The LMD-X2705MD/X2700MD is intended for use in the electromagnetic environment specified below. The customer or the user of the LMD-X2705MD/X2700MD should assure that it is used in such an environment. Compliance level IEC 60601 Immunity test Electromagnetic environment –...
  • Page 6 Guidance and manufacturer’s declaration – electromagnetic immunity The LMD-X2705MD/X2700MD is intended for use in the electromagnetic environment specified below. The customer or the user of the LMD-X2705MD/X2700MD should assure that it is used in such an environment. IEC 60601 Immunity test Compliance level Electromagnetic environment –...
  • Page 7 To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the LMD-X2705MD/X2700MD is used exceeds the applicable RF compliance level above, the LMD-X2705MD/X2700MD should be observed to verify normal operation.
  • Page 8 The LMD-X2705MD/X2700MD is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. Portable RF communications equipment should be used no closer than 30 cm (12 inches) to any part of the LMD-X2705MD/X2700MD. Otherwise, degradation of the performance of this equipment could result. IEC 60601...
  • Page 9 – Top side: 30 cm (11 in.) or more Warning on power connections for Consult with Sony qualified personnel for the following medical use types of installation location. Customers in the U.S.A. and Canada should use the –...
  • Page 10 Disposal of Old Electrical & Electronic Equipment For the customers in the U.S.A. (Applicable in Republic of India) SONY LIMITED WARRANTY - Please visit http:// www.sony.com/psa/warranty for important information and complete terms and conditions of Sony’s limited warranty applicable to this product.
  • Page 11: Table Of Contents

    Specifications ..........30 Table of Contents Dimensions ............36 Licenses ............37 Precaution ............12 On Safety ...........12 On Installation ...........12 Precautions for using this unit safely ..12 Precautions for connecting this unit with other medical devices ......12 To prolong the life of the unit ....12 Cautions for RESPONSIBLE ORGANIZATION when connecting this equipment to IT- NETWORK ..........12...
  • Page 12: Precaution

    Some people may experience discomfort (such NETWORK as eye strain, fatigue, or nausea) while watching video images. Sony recommends – connection of the PEMS to an IT-NETWORK that that all viewers take regular breaks while includes other equipment could result in watching video images.
  • Page 13: On Simultaneous Use With An Electrosurgical Knife, Etc

    changes in the IT-NETWORK configuration; the physical characteristics of the liquid crystal  connection of additional items to the IT- display, such “stuck” pixels may appear  NETWORK; spontaneously. These problems are not a disconnecting items from the IT-NETWORK; malfunction. ...
  • Page 14: On Fan Error

    A material that withstands disinfection is used fan error indication appears on the screen, turn for the front protection plate of the medical use off the power and contact an authorized Sony LCD monitor. The protection plate surface is dealer.
  • Page 15: On Repacking

    U.S.A., Canada and Europe have been obtained LMD-X2700MD  for this monitor. The illustration of LMD-X2705MD is used for the The monitor is designed for use in the medical explanations. Any differences in specifications treatment field, with the sheet switch, screen are clearly indicated in the text.
  • Page 16: Location And Function Of Parts And Controls

    Location and Function of Parts and Controls Front Panel   (On)/ (Standby) switch /// buttons Press the  side to turn the monitor on. Press the Press to select the items and setting values. side to switch the unit to standby state. ...
  • Page 17: Input Signals And Adjustable/Setting Items

     Input select buttons connector assigned to the PORT A/B is PORT A/PORT B: Each input connector can displayed. be assigned for PORT A/B. When you press PORT A or PORT B when Display Port 1 is assigned for PORT A/B in the it is lit in white, an image from the input default setting.
  • Page 18: Rear/Bottom Panel

    Rear/Bottom Panel  HDMI cable holder Note Secures the HDMI cable (Ø7 mm or less). Consult with Sony qualified personnel at using this terminal. Close Caution For safety, do not connect the connector to peripheral device wiring that might have excessive voltage.
  • Page 19 If use of such a device is unavoidable, isolate its power supply by connecting an isolation  3G/HD/SD-SDI input/output connector transformer, or by connecting an isolator (BNC type) (LMD-X2705MD only) between the connecting cables. input connector After implementing these measures, confirm...
  • Page 20: Preparation

    Shutter Preparation AC input connector Connecting Make sure that the  (On)/ (Standby) switch is set to (Standby). AC power cord AC plug holder Remove the connector cover. While pressing the lock levers to the direction , slide off the connector cover. When using the AC adaptor (optional) Slide the shutter to show the DC input connector and insert the DC connector to the...
  • Page 21: Turning On The Monitor/Switching Input Settings

    Initial Setting When you turn on the unit for the first time after purchasing it, select the area and language where you intend to use this unit from among the options. To set the using area To unplug the AC power cord Turn on the unit.
  • Page 22: Using The Menu

    To set the using language Using the Menu You can select one of eight languages (English, Chinese, Japanese, Italian, Spanish, German, French, and Russian) for display on the menu and The unit is equipped with an on-screen menu for other on-screen displays. The default menu making various adjustments and settings such language is set to “English.”...
  • Page 23: Adjustment Using The Menus

    Press the / button to select the tab. Adjustment Using the Menus The selected tab is shown in blue, and setting items of the selected tab appear. Items Select an item. Press the / button to select the item. The screen menu of this monitor consists of the The item to be changed is displayed in blue.
  • Page 24: Adjusting And Changing The Settings

    OSD Setting Submenu Setting Power On Setting Color Temperature Select the color temperature from Power Save among “D56,” “D65,” “D93.” Serial Remote Remote Note Ethernet Setting If the setting is changed, Gain R/G/B Custom Button Offset and Bias R/G/B Offset are Panel Display restored to 0 respectively.
  • Page 25: Pip / Pop Menu

    Room Camera Surgical Camera with “Enhanced Format,” use a  Microscope Premium High-Speed HDMI cable  within a length of 3 meters (Sony  Vital Device product recommended). Power Supply  5V Output (DVI-D): Select “On” when the 5 V output power of the DVI-D input connector is output.
  • Page 26: System Configuration Menu

    Submenu Setting Submenu Setting Auto Input Select Select the auto detection setting of the  Off: input signal from “Mode1,” “Mode2,” The Auto Input Select function does “Off.” not operate.  Mode1: Inputting signals to the input System Configuration menu connector automatically changes connectors.
  • Page 27 Submenu Setting Submenu Setting  Auto: Power Save The format and scan mode are Energy Saving You can select the energy saving mode displayed temporarily when the Mode from the following options. content of Status Display is  Off: changed. Turns the energy saving mode off.
  • Page 28: Initial Setup Menu

    Set to “Not Skip” or “Skip” for the Display Port 1, Display Port 2, HDMI, Backlight Adjusts the brightness of the display. A DVI-D, SDI (LMD-X2705MD only) higher setting increases the brightness connector. of the display, and a lower setting darkens the display.
  • Page 29: Preset Menu

    Submenu Setting Troubleshooting Auto Input Select  Mode1: Inputting signals to the input connector automatically changes connectors. This section may help you isolate the cause of a  Mode2: problem and as a result, eliminate the need to Gives priority to PORT A and switches the PORT automatically.
  • Page 30: Error Messages

    HDMI connector (1) HDMI 2.0b, HDCP1.4 correspondence DVI-D input connector DVI-D connector (1) TMDS single link 3G/HD/SD-SDI input connector (LMD-X2705MD only) BNC type (1) SD: SMPTE ST 259 compliant HD: SMPTE ST 292-1 compliant 3G: SMPTE ST 424 compliant Remote connector...
  • Page 31 CD-ROM (including the WARRANTY, OR FOR ANY OTHER REASON Instructions for Use) (1) WHATSOEVER. Service Contact List (1) SONY WILL NOT BE LIABLE FOR CLAIMS OF  Information for Customers in ANY KIND MADE BY USERS OF THIS UNIT OR Europe (1) MADE BY THIRD PARTIES.
  • Page 32 Pin number Signal REMOTE (stereo mini jack) connector Pin number Signal On (short to sleeve) Off (open) Insert detection 5V 2.0A (DC output) connector Pin number Signal 12V 2.5A (DC output) connector Pin number Signal 12 V...
  • Page 33 Available signal formats The unit is compatible with the signal systems shown below: Signal format (LMD-X2705MD only) SD-SDI  720×487/59.94i 4:2:2 YCbCr 10bit  720×576/50i 4:2:2 YCbCr 10bit HD-SDI  1920×1080/24p 4:2:2 YCbCr 10bit  1920×1080/25p 4:2:2 YCbCr 10bit ...
  • Page 34 Signal format Display Port 1, 2 HDMI DVI-D    4:4:4 RGB 10bit/8bit   × 4:4:4 YCbCr 10bit/8bit 1920×1080/60p  × × 4:2:2 YCbCr 12bit  × × 4:2:2 YCbCr 10bit/8bit   × 4:4:4 RGB 8bit  ...
  • Page 35 PC signal (DVI) Range of DVI input signal (Compatible with up to 1920 × 1080/60 Hz) Vertical frequency: 50.0 Hz to 85.1 Hz Horizontal frequency: 31.0 kHz to 75.0 kHz Dot clock: 25.175 MHz to 148.5 MHz Picture size, phase: automatic discrimination by the DE (Data Enable) signal Displays normally up to a maximum horizontal resolution of 1920 dots.
  • Page 36: Dimensions

    Side Dimensions When an optional stand SU-600MD is attached Front When an optional stand SU-600MD is attached Unit: mm (inches) Mass: Approx. 8.8 kg (19 lb 6.4 oz) (when the optional stand is not installed) Rear (VESA Mount Instruction) * Length of M4 screws (4) Monitor cabinet Attached object M4 screw...
  • Page 37: Licenses

    Each condition informs users that they have the rights to procure, alter, and redistribute the software source code. For details on the source code, contact your local Sony representative. However, be aware that Sony will not answer questions concerning the data in the source code.