2-18
Calibration

Remote calibration procedure

Step 1: Prepare the Model 2430 for calibration
1.
2.
3.
Step 2: Voltage calibration
1.
2.
3.
Connect the Model 2430 to the controller IEEE-488 interface or RS-232 port using a
shielded interface cable.
Turn on the Model 2430 and the test equipment, and allow them to warm up for at least
one hour before performing calibration.
If you are using the IEEE-488 interface, make sure the primary address of the Model
2430 is the same as the address specified in the program you will be using to send com-
mands. (Use the MENU key and the COMMUNICATION menu to access the IEEE-488
address.)
Connect the Model 2430 to the digital multimeter, and select the multimeter DC volts
function. (See Figure 1-2).
Send the commands summarized in Table 2-8 in the order listed to initialize voltage cal-
ibration. (When the :CAL:PROT:CODE command is sent, the instrument will assume
the operating states listed in Table 2-2.)
Perform the range calibration steps listed in Table 2-9 for all ranges. For each range:
• Send the :SOUR:VOLT:RANG command to select the source and sense range being
calibrated. For example, for the 2V range, the following command would be sent
:SOUR:VOLT:RANG 2
• Program the source to output the negative full-range value using the :SOUR:VOLT
command. For example:
:SOUR:VOLT -2
• Note and record the multimeter reading.
• Use the multimeter reading as the parameter for the :CAL:PROT:SOUR and
:CAL:PROT:SENS commands. For example, a typical value for the 2V range would
be:
:CAL:PROT:SOUR -1.998
:CAL:PROT:SENS -1.998
• Program the voltage source for 0V output using the :SOUR:VOLT 0.0 command.
• Note the multimeter reading.
• Send the source and sense calibration commands using the multimeter reading for the
parameter. For example:
:CAL:PROT:SOUR1E-3
:CAL:PROT:SENS 1E-3
• Set the source to the positive full-range value using the :SOUR:VOLT command. For
example:
:SOUR:VOLT 2
• Note and record the multimeter reading.