2-14
Calibration
Table 2-4
Front panel current calibration
Source range
1. Use EDIT and RANGE keys to select source range.
2. Multimeter reading used in corresponding calibration step. See procedure.
3. Current calculated as follows: I = V/R, where V is the DMM reading, and R is the
characterized value of the 1Ω resistor. 10A pulse range calibrated simultaneously.
Step 4: Enter calibration dates and save calibration
NOTE
1.
2.
1
Source current
10μA
+10.000μA
+00.000μA
-10.000μA
+00.000μA
100μA
+100.00μA
+000.00μA
-100.00μA
+000.00μA
1mA
+1.0000mA
+0.0000mA
-1.0000mA
+0.0000mA
10mA
+10.000mA
+00.000mA
-10.000mA
+00.000mA
100mA
+100.00mA
+000.00mA
-100.00mA
+000.00mA
1A
+1.0000A
+0.0000A
-1.0000A
+0.0000A
3
3A
+3.0000A
+0.0000A
-3.0000A
+0.0000A
For temporary calibration without saving new calibration constants, proceed to
Step 5: Lock out calibration.
From normal display, press MENU.
Select CAL, and then press ENTER. The Model 2430 will display the following:
CALIBRATION
UNLOCK EXECUTE VIEW-DATES
SAVE LOCK CHANGE-PASSWORD
ß
Multimeter
current reading
___________ μA
___________ μA
___________ μA
___________ μA
___________ μA
___________ μA
___________ μA
___________ μA
___________ mA
___________ mA
___________ mA
___________ mA
___________ mA
___________ mA
___________ mA
___________ mA
___________ mA
___________ mA
___________ mA
___________ mA
___________ A
___________ A
___________ A
___________ A
___________ A
___________ A
___________ A
___________ A
©
2