Generating An Srq On Calibration Complete - Keithley 2430 Service Manual

1kw pulse sourcemeter
Hide thumbs Also See for 2430:
Table of Contents

Advertisement

Generating an SRQ on calibration complete

An IEEE-488 bus SRQ (service request) can be used to detect operation complete instead of
repeatedly polling the Model 2430. To use this method, send both *ESE 1 and *SRE 32 to the
instrument, then include the *OPC command at the end of each calibration command line, as
covered above. Clear the SRQ by querying the ESR (using the *ESR? query) to clear OPC sta-
tus, then request the status byte with the *STB? query.
Refer to your controller's documentation for information on detecting and servicing SRQs.
Command Reference
B-11