10
2.3

Diagnostic level

With the different functions in the diagnostic level you can interrogate and simulate operating con-
ditions of the unit. These functions may help you to check the correct function of the unit and its
components.
When leaving the diagnostic level modified settings are reset.
2.3.1

Selecting the diagnostic level

Call the diagnostic level from the service level (see overview in chapter 2.1).
2.3.2

Operating status of the carriage/rear cover lock switch

(see chapter 2.3.15)
Note: This is a display function only (no function for PH27).
Operating status:
2.3.3

Operating status of the leakage sensor (option)

(see chapter 2.3.2)
Leakage sensor
(see chapter 2.3.4)
Note: This is a display function only.
Operating status:
Cover
Closed
(see chapter 2.3.3)
Closed (carriage/rear cover inserted and locked)
Open (carriage/rear cover not inserted or locked, lock switch is open, error
1 should be displayed).
Closed
Closed (contact of the leakage sensor is closed, no leakage present)
Open (leakage sensor has triggered or is defective, error 5 should be dis-
played)