3.2 Configuration Set Up Procedure

ATTENTION:
In Table 3-2, under the Press column:
• Hold means to hold the button down until the display changes.
• Momentarily means to press and release the indicated button.
From the Online pH display, follow this procedure.
ATTENTION:
If no key is pressed for 60 seconds, the display will abort the entry mode and default to Online Display.
Step
Operation
Enter Buffer Group
1
Selection
Edit Buffer Group
Select desired Buffer
Group
Save the Buffer Group
Enter Solution
2
Temperature
Coefficient Selection
Edit Solution
Temperature Coefficient
Select desired
Coefficient
Save the Solution
Temperature coefficient
2/04

Table 3-2 Configuration Set Up Procedure

Press
bFrG (for 1 second) then,
MODE
Hold
(Current Buffer Group Selection)
MODE
Flashing Display – You are now in EDIT mode
Hold
(Value of current Buffer Group selection)
To select US (default), NIST, or Euro
Momentarily
MODE
Selection for group
Momentarily
COEF (for 1 second) then,
MODE
Momentarily
(Solution Temperature Coefficient Selection)
MODE
Flashing Display – You are now in EDIT mode
Hold
To select:
Momentarily
Selection for coefficient
MODE
Momentarily
DirectLine® DL421 Sensor Module User Manual
Display
(Value of current Coefficient selection)
0.00 pH/10°C (default) or
−0.16 pH/10°C (pure water)
−0.32 pH/10°C (AVT, Amine,
Phosphate or Oxygenated Treatment
Configuration
15