2.2 Integral Mounting

Table 2-2 Integral Mounting Procedure (refer to Figure 2-3)

Step
Connect Electrode to Pipe and Electronics Assembly
1
Screw the electrode into the pipe tee (3/4 " NPT thread).
Make sure that the final position of the installed electronics module allows the display to
be easily viewed by plant personnel.
2
Align the slots in the electronics module with those in the electrode and press down to
connect the electronics to the electrode.
3
Tighten the locking screw on the bottom rear of the electronics module.
2/04
Step 2

Figure 2-3 Integral Mounting