Configuration Set Up Procedure; Table 3-1 Configuration Set Up Procedure - Honeywell DirectLine DL423 User Manual

Sensor module for conductivity/resistivity measurements
Table of Contents

Advertisement

Configuration

3.2 Configuration Set Up Procedure

ATTENTION:
In Table 3-1, under the Press column:
 Hold means to hold the button down until the display changes.
Momentarily means to press and release the indicated button.
From the Online display, follow this procedure.
ATTENTION:
If no key is pressed for 60 seconds, the display will abort the entry mode and default to
Online Display.
Step
Operation
1
Enter Cell
Constant Selection
(if applicable)
Edit Cell Constant
Select desired Cell
Constant
Save the Cell
Constant
2
Enter PV Type
Selection
Edit PV type
Selection
Select desired PV
Type
(selection
determined by cell
constant)
Save the PV Type
continued
14

Table 3-1 Configuration Set Up Procedure

Press
MODE
Hold
MODE
Hold
Momentarily
MODE
Momentarily
MODE
Momentarily
MODE
Hold
Momentarily
MODE
Momentarily
DirectLine® DL423 Sensor Module User Manual
Display
CnSt (for 1 second) then,
(Current Cell Constant Selection)
Flashing Display – You are now in EDIT
mode
(Cell Constant Value)
To select 0.01, 0.1(default), 1, 10, 25, 50
Selection for Cell Constant
PtYP (for 1 second) then,
(Current Selection)
Flashing Display – You are now in EDIT
mode
(Value of current PV Type selection)
To select:
S, mS (default), rES, PPb, PPm, PPt,
Conc
Selection for PV Type
9/11