Specifications - Model 5100 - AEMC 5100 User Manual

5000v megohmmeter
Table of Contents

Advertisement

Megohmmeter Models 5000N/5100/5110

Specifications - Model 5100

INSULATION TESTS
DC Test Voltages:
500, 1000, 2500, 5000V
Megohm Ranges:
For each test voltage two direct reading ranges:
30 to 30,000MΩ
3 to 3000GΩ
(3000 to 3,000,000MΩ)
Short Circuit Current: 6mA (max)
Accuracy: 5% of reading typical
Charging Time:
0.5 to 5 seconds/µF typical
Discharging Time:
Automatic discharge when test button is released; 0.1 seconds/µF.
Discharge voltage displayed on meter.
Test Voltage Generation:
Solid state circuitry generating rated test voltage across the full range
RESISTANCE TESTS
Test Current: Constant 33 µA DC
Kilohm Range: 10 to 30,000kΩ (30MΩ)
Maximum Test Voltage: 1300V
DC
Accuracy: 5% of reading typical
VOLTAGE TESTS (SAFETY CHECK)
Voltage Range: 0 to 600V
/
AC
DC
Accuracy: 3% of full scale
- 13 -

Advertisement

Table of Contents
loading

This manual is also suitable for:

51105000n

Table of Contents