Megohmmeter Models 5000N/5100/5110
GENERAL SPECIFICATIONS
Audible Test Signal:
Ten beeps per minute approx. (signal may be disabled)
Power Supply:
Rechargeable 12V NiCad battery
Typical life: 500 to 2500 1 min tests (depending on test voltage)
Recharging time: 14 hrs (max) for full charge
Fuse protection: 3.15A fast blow
Low battery indicators
AC Supply: 110/220V, 47 to 450Hz AC ± 20%
Dielectric Test: 4000V
, 60Hz, 1 minute
AC
Overload: 600Vrms between one terminal and the other two
Meter Movement/Display:
Rugged taut band suspension;
4.4 x 2.2" (112 x 55mm) meter with 4 scales
Clock/Timer: Test voltage ON activates timer automatically
Dimensions: 9.8 x 10.2 x 15.4" (250 x 260 x 390mm)
Weight: 14.8 lbs (6.7kg)
Temperature Range: 14° to 131°F (-10 to 55°C)
Field Case:
High impact, fiberglass charged polycarbonate (fire resistant UL94);
sealed and weatherproofed; rainproof when closed
Terminals:
Color-coded safety terminals; guard terminal minimizes surface
leakage errors
Safety Standards:
4000V dielectric test
Double insulation
IEC 348
Electromagnetic Compatibility:
CEI 801-2: Electrical Discharge. Class II
CEI 801-3: Magnetic Fields 3V/m
CEI 801-4: Rapid Transients class II
CEI 801-5: Electrical shock (5KV 2 Joules)