Ac Supply Module - AEMC 5100 User Manual

5000v megohmmeter
Table of Contents

Advertisement

Megohmmeter Models 5000N/5100/5110
7. Charging Light:
Indicates the battery is charging or operation from AC supply
8. Power Line Socket
9. Fuse Socket:
Select 110/220V with small connector in this socket.
NOTE: When switching the power supply between 110V and 220V,
the selection module must be changed accordingly. Change fuse
before making any connections.
10. Analog Measurement Scale (Model 5110 has color scale)
11. Zero Adjust
12. Reset Button:
Resets chronometer and selects time values during programming.
13. Clock/Timer:
Toggles between the clock and timer on the display.
14. LCD Display: Clock or timer.
15. Selection Switch

AC Supply Module

Models 5100 and 5110 may be powered by 110V or 220V
450Hz). The instrument includes a 110V supply cord, which provides
power to the instrument as well as charging voltage for the rechargeable
battery. Verify that "110V" appears in the power supply module window
opening; this will permit operation with the 110V supply cord.
To change the power supply voltage selection, push the tab on the
selection module to the right and remove the module. Remove the
portion of the module that contains the fuse assembly and rotate 180° to
display "220V" in the module window. A 220V supply cord (not supplied)
is required for 220V operation.
CAUTION: Do not change the power supply voltage while operating
either Model 5100 or Model 5110 from the AC supply.
Both Models are protected by a 3.15A, 250V fuse (Cat. #1007.26).