Disassembly - AEMC 5100 User Manual

5000v megohmmeter
Table of Contents

Advertisement

Disassembly

Detachable Cover
The cover hinges are fitted with
spring loaded clips which allow
the cover to be removed. To
detach
the
cover
Models 5100 or 5110, open lid to
a horizontal position and apply
downward pressure to the hinge
side of the cover while gripping it
firmly.
To re-attach the cover, position
the cover in a horizontal position
and fit the hinges into the
respective housing. Apply strong
rear-to-front
pressure
cover until it snaps into place.
Figure 4
from
the
to
the
Megohmmeter Models 5000N/5100/5110
Chassis
Ribbon Cable
Case
Fastening Screws
- 19 -
Use the hex key to
unscrew
the
fastening screws from
the chassis, which are
located on the bottom
of the case.
Pull out the chassis.
Do not forget to unplug
the ribbon cable which
connects
the
supply board in the
bottom of the case to
the boards which are
mounted
chassis.
Unhinging
Force
Supporting Point
Figure 5
six
power
in
the
Force

Advertisement