Leica TS12P User Manual page 49

Robotic total station
Table of Contents

Advertisement

Next step
TS12P, Check & Adjust
Step
Description
7.
TOOLS Adjustment Accuracy
<No.of Meas:> Shows the number of runs executed. One run consists of a
measurement in face I and face II.
< l Comp:> and similar lines show the standard deviations of the deter-
mined adjustment errors. The standard deviations can be calculated from
the second run onwards.
It is recommended to measure at least two runs.
8.
MEAS (F5) if more runs have to be added. Continue with step 3.
OR
CONT (F1) to accept the measurements and to proceed to TOOLS Adjust-
ment Results. No more runs can be added later.
IF the results are
THEN
to be stored
CONT (F1) overwrites the old adjustment errors with the new
ones, if the Use status is set to Yes.
to be determined
REDO (F2) rejects all new determined adjustment errors and
again
repeats the whole procedure. Refer to step 3. of paragraph
"Combined adjustment procedure step by step".
49