Peristaltic Pump Motor

When required
Tools required
Parts required

Removing the peristaltic pump motor

1 In the user interface start the maintenance mode and select the "Change
2 Switch off the sampler at the main power switch. Remove the power cable.
3 Remove the top cover, top plate, and foam (see
4 Remove the plate tray from the tray base.
5 Remove the transport assembly,
6 Remove the sampling unit, (see
7 Remove the air channel.
8 Remove the peristaltic pump, (see
9 Remove the base plate of the pump with a small flat-head screwdriver.
10 Disconnect the peristaltic pump motor from the SUD board (labelled
11 Take out the motor and the holder.
12 Unscrew the 2 screws which hold the peristaltic pump motor.
13 Remove the peristaltic pump motor.
1100 Series WS MWS Reference Manual
When defective.
1/4 inch-5/16 inch wrench 8710-0510 (supplied in accessory kit)
4 mm open end wrench 8710-1534 (supplied in accessory kit)
Pozidrive screwdriver
Flat-head screwdriver (if interface board installed)
Peristaltic pump motor, 5065-4409
Needle/Seat" function.
page 132).
page 133).
"mixing M").
Repairing the Sampler
"Top Cover and Foam"
(see"Removing the sample transport"
"Removing the sampling unit"
"Peristaltic Pump"
on
on
on page 134).
on page 128).
137