Repairing the Sampler

Installing the metering drive motor and belt

1 Feed the motor cable through to the SUD board. Connect the cable to
2 Install the motor with the four fixing screws.
3 Ensure the belt is seated correctly over the gear and motor shaft.
4 Install the sampling unit, (see
5 Install the sample transport, (see
6 Install the top cover, top plate, and foam (see
7 Replace the plate tray in the tray base.
8 Turn ON the sampler, close the front door.
9 The system will reset automatically.
10 In the user interface perform a "Sample Transport Self Alignment".
142