RDO (Optical Dissolved Oxygen Sensor) Specifications
Accuracy
Range
Resolution
Sensor Type
Response
Time
Units of
Measure
Methodology
ORP Sensor Specifications
Accuracy
Range
Resolution
Sensor Type
Response
Time
Units of
Measure
Methodology
800-446-7488
±0.1 mg/L; ±0.2 mg/L; ±10% of reading
0 to 8 mg/L; 8 to 20 mg/L; 20 to 50 mg/L; Full operating
range: 0 to 50 mg/L
0.01 mg/L
Fixed with replaceable RDO Sensor Cap (life: 1 year
typical)
T90: <45 sec. T95: <60 sec.
mg/L, % saturation, ppm
EPA-approved In-Situ Methods 1002-8-2009 1003-8-2009
1004-8-2009
±5.0 mV
±1400 mV
0.1 mV
Replaceable pH/ORP combo sensor
<15 sec.
mV
Std. Methods 2580
8