Cleaning the Optical Unit

The only maintenance action which you must perform is checking the image
quality. Refer to the User Manual of the NX
To clean the optic unit, proceed as follows:
1. Remove the cassette unit.
2. Move the cleaning lever from left to right and back.
This is the location of the cleaning lever:
CAUTION:
Dust can cause stripes in the image, parallel to the image
plate movement.
When you recognize this type of artefact, clean the optic unit
using the cleaning brush.
TM
software.