80 | CR 10-X, CR 12-X, CR 15-X | Remarks for HF-emission and immunity
RF emissions in accordance
with CISPR 11
Harmonic emissions in
accordance with IEC
61000-3-2
Voltage fluctuations /
flicker emissions in
accordance with
IEC 61000-3-3
This device is intended for operation in the electromagnetic environment
given below. The user of the device should ensure that it is used in such an
environment.
Resistance to
Jamming
Discharge of
static
electricity in
accordance
with IEC
61000-4-2
Fast transient
electrical
disturbance
variables /
bursts in
accordance
with IEC
61000-4-4
Impulse
voltages
(surges) in
accordance
with IEC
61000-4-5
2491E EN 20150409 1543
IEC 60601 Test
Level
Test
± 6 kV contact
discharge
± 8 kV air
discharge
± 2 kV for
network leads
± 1 kV for entry
and outlet leads
± 1 kV push-pull
voltage
± 2 kV common
mode voltage
Class A
The device is suitable for use in
all establishments other than
domestic, and may be used in
Class A
domestic establishments and
those directly connected to the
public low-voltage power supply
network that supplies buildings
Complies
used for domestic purposes,
provided the following warning
is heeded.
Level of
Agreement
± 6 kV contact
discharge
± 8 kV air
discharge
± 2 kV for
network leads
± 1 kV for entry
and outlet leads
± 1 kV push-pull
voltage
± 2 kV common
mode voltage
Electromagnetic
Environment
Guidelines
Floors should consist
of wood, concrete or
ceramic tiles. The
relative humidity
must be at least 30%,
if the floor is made of
synthetic material.
The quality of the
voltage supplied
should correspond to
a typical commercial
or clinical
environment.
The quality of the
voltage supplied
should correspond to
that of a typical
commercial or clinical
environment.