24 | CR 10-X, CR 12-X, CR 15-X | Introduction

Installation

The digitizer is equipped with handles at the bottom left and right sides to
move the device easily to another location.
For two persons to lift the digitizer, each should stand at the side of the
digitizer and hold it by the handle with both hands.
For one person to lift the digitizer, remove the cassette unit to reduce the
weight, stand in front of the digitizer and hold it by the handles.
2491E EN 20150409 1543
WARNING:
When using the power supply, care must be taken to ensure that
there is either a mains plug or an all-cable disconnecting device
in the internal installation fitted near the device and that it is
easily accessible in case of emergency.
CAUTION:
Do not lift the device by holding the cassette unit or by holding
the back cover.
CAUTION:
The digitizer and the cassette storage shall be protected against
direct radiation in such a way, that the annual dose equivalent
at the place of installation will not exceed 1 mSv/a.
CAUTION:
If the digitizer is installed inside an X-ray room it must be
protected from stray radiation by proper shielding.
CAUTION:
The device is a table-top digitizer. The structure and stability of
the table used need to be suitable in relation with the size and
weight of the system. Do not use excessive force when inserting
cassettes in the digitizer as the device may slip or drop off the
table. Use a non-slip-mat below the digitizer or other anti-skid
measures. The table should not be subject to excessive shock
and vibrations from other sources, as this may disturb the
operation of the digitizer.