CR 10-X, CR 12-X, CR 15-X | Operating CR 10-X, CR 12-X and CR 15-X | 61

Connection Problems

In case the status indicator of the digitizer is blinking red, the user should look
at the "status" of the Digitizer Remote Display to decide whether digitizer
internal problems or connection problems occurred.
If an error message is displayed on the NX PC, the user is informed which
actions to perform to solve the problem.
In case no error message is displayed on the screen, a connection problem
occurred.
Condition
Connection
problem
between
digitizer and
the Digitizer
Remote
Display.
Connection
problem
between
digitizer and
NX PC.
CAUTION:
Operation failure of the device can cause delayed diagnosis.
Check if Digitizer Remote Display is running.
Message at
Digitizer
Remote
Display
No error
message on NX
PC.
Status
Action
indicator
Red blinking
Check if Digitizer
fast
Remote Display is
running. Start/restart
Digitizer Remote
Display.
Red blinking -
Check the Ethernet
3 pulses
cables. If the error
remains, restart PC and
digitizer or call service.
2491E EN 20150409 1543