CR 10-X, CR 12-X, CR 15-X | Operating CR 10-X, CR 12-X and CR 15-X | 65

Image Plate Transport Problems

Details
Possible
causes
Solutions
The following error message is displayed in the Digitizer
Remote Display:
Image plate is not erased! Remove the cassette and press the
confirm button. Do not use the cassette again before an
inspection.
Image plate is not scanned and erased! Remove the cassette
and press the confirm button. Do not use the cassette again
before an inspection.
A problem occurred when transporting the image plate in the
digitizer.
Remove the cassette and press the confirm button and
perform following checks:
1. Check the cassette for damages.
2. Open the cassette and check the shutter for damages.
3. Check if image plate is bent. Put the image plate on a flat
surface. The entire image plate must touch the surface.
The image plate must be exchanged if there is a gap
between surface and a part of the image plate e.g. if you
can see a lifted edge or a sharp bend.