CR 10-X, CR 12-X, CR 15-X | Remarks for HF-emission and immunity | 83
investigation of the location is recommended, to ascertain the
electromagnetic environment as a result of stationary high frequency
transmitters. If the field strength of the device exceeds the level of
agreement given above, the device must be observed with regard to its
normal operation at each place of use. In case of unusual performance
characteristics, it can be necessary to take additional measures, such as the
re-orientation of the device, for example.
The field strength will be lower than 3 V/m above the frequency range from
150 kHz to 80 MHz.
• REMARK 1: The higher value will apply at 80 MHz and 800 MHz.
• REMARK 2: These Guidelines may not apply to all situations. The
dispersion of electromagnetic waves is influenced by absorption and
reflections from buildings, objects and people.
Radio frequency communications equipment can effect medical electrical
equipment.
This device is intended for operation in an electromagnetic environment in
which the radiated high frequency disturbance variables are monitored. The
user of the device can help to prevent electromagnetic disruptions by
maintaining the minimum distances between portable and mobile high
frequency communication equipment (transmitters) and the device as
recommended below, in accordance with the maximum output power of the
communications equipment.
Recommended Protective Distances between Portable and Mobile High
Frequency Communication Equipment and the Device
Rated Power
of the
Transmitter
W
150 kHz to 80
MHz
d = 1.2
0.01
0.12
0.1
0.38
1
1.2
Protective Distance in accordance with Transmission
80 MHz to 800
MHz
d = 1.2
0.12
0.38
1.2
Frequency
m
800 MHz to 2.5
GHz
d = 2.3
0.23
0.73
2.3
2491E EN 20150409 1543