Cannot read data on the image plate

Details
Possible
causes
Solutions
Related Links
Cleaning the Optical Unit
2491E EN 20150409 1543
The following error message is displayed in the Digitizer
Remote Display:
Error during read out of data on the image plate. Remove the
cassette and press the confirm button. Do not use the
cassette again before an inspection.
Defect/soiled bar code on image plate
Optics cleaning lever in optical path and not positioned at
left side.
Remove the plate from the cassette -as described in the AGFA
CR Plates and Cassettes User Manual- and check if the
barcode is fully legible. If necessary, remove any dirt
(conform the plate cleaning instructions).
At the left side, put the optics cleaning lever back to the
"home" position so that the digitizer can read the barcode on
the image plate.