Figure 4-3: Calibration Screen; Figure 4-4: Cal Screen - 3M SE-401 User Manual

Se-400 series sound level meter
Hide thumbs Also See for SE-401:
Table of Contents

Advertisement

18
Calibrate
Performing a Calibration
6. In the calibration screen, either press the CAL softkey or press the
 Note: the calibration screen details the last calibration and is saved into the memory of the instrument.
If you download your data to DMS (software), the calibration history is saved to the session and is
viewable/printable in the software.

Figure 4-3: Calibration screen

7. In the calibration screen, press the corresponding
specific dB level (noted on the calibrator).

Figure 4-4: CAL screen

8. Press the
9. Press the
key to save the changes or escape to abort.
key twice to return to the main measurement screen.
key.
Opening CAL screen (to perform calibration)
Either press the
key or press the Cal softkey
softkeys or press
Calibrating
Match the dB level from the calibrator to the SE-
400 Series by pressing either the corresponding
softkeys or the up/down arrow keys
Sound Examiner SE-400 Series
/
to adjust to the

Hide quick links:

Advertisement

Table of Contents
loading