Leak Tests - Dräger FPS 7000 S-fix Instructions For Use Manual

Full face masks
Hide thumbs Also See for FPS 7000 S-fix:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 19
Maintenance
1. Remove accessories and clean separately
2. Clean all parts with lukewarm water with the addition of Sekusept
Cleaner and clean with a soft cloth (temperature: max. 30 °C,
2)
concentration: 0.5 - 1 %)
.
3. Rinse all parts thoroughly under running water.
4. Prepare a disinfection bath consisting of water and Incidin
(temperature: max. 30 °C, concentration: 1.5 %)
5. Place all parts that require disinfecting into the disinfection bath
(duration: 15 minutes).
6. Rinse all parts thoroughly under running water.
7. Allow all parts to dry in the open air or in a drying closet (temperature:
max. 60 °C). Protect against direct sunlight.
8. Fit the accessories.
Relubricate the sliding surfaces (ball joint, guide
Mask/helmet
plate) of the mask/helmet adapter with Voltalef
combination
®
Molykote
111.
9. Apply a thin layer of Vaseline or Molykote
piece (figure A3-18 on page 3) in the area of the sealing surfaces.
NOTICE
i
i
The facepiece is also suitable for mechanical cleaning and
disinfection. Further information can be obtained from Dräger.
1)
Follow the relevant instructions for use.
®
2)
Sekusept
is a registered trademark of Ecolab Deutschland GmbH. Follow
the relevant information for the product.
®
3)
Incidin
is a registered trademark of Ecolab USA Inc. Follow the relevant
information for the product.
28
1)
.
4.3

Leak tests

®
The leak test should be performed with an appropriate test device
(e. g. the Testor or Quaestor series).
NOTICE
®
Rapid
i
i
3)
.
When testing facepieces sized L or S, place the mask care-
fully on the test heads sized M, since the face seal is not op-
timally sealed at times. Moistening the face seal facilitates
exact donning because the mask can slide better on the test
heads.
(1) Checking leak tightness
1. Moisten the face seal and exhalation valve with water.
2. Mount the facepiece on the test head, use a holder suitable for
®
mask/helmet combinations, if necessary.
or
3. Make sure that the face seal fits tightly onto all parts of the test
head. Inflate the test head, if necessary.
®
111 to the connection
4. Adjust the connector with adapter tightly and produce a negative
pressure of 10 mbar.
The facepiece is considered leak-tight if the pressure drop is less than
1 mbar per minute. If the facepiece is leaky, continue with (2).
(2) Test with sealed exhalation valve
1. Remove exhalation valve.
2. Plug in the exhalation valve test plug.
3. Generate a negative pressure of 10 mbar.
4. If the pressure drop is less than 1 mbar/min, remove the exhalation
valve test plug and insert a new valve disc.
5. Repeat the leak test (1). If the facepiece is leaky, continue with (3).
(Certified by Dekra Exam)
Dräger FPS 7000

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Fps 7000Fps 7000 q-fix

Table of Contents