Lauda Integral XT 150 Operating Instructions Manual page 66

Integral xt series. process thermostats / high-temperature thermostats
Table of Contents

Advertisement

or
Wait 4 seconds or
12!45
Command
Control Variable
Pump
Menu
64
Change temperature probe source:
1
-
EP
-
EA
-
ES
-
receiving actual values from a PC.
-
Display flashes 4 s à new value is automatically accepted, or
-
Value is immediately accepted with Enter key.
-
If
°C
then the green
the external temperature signal.
Internal
External Pt100
Analog module
Digital module
End
T
set
for the internal probe,
only when an external probe is connected,
only when an analog module is connected and configured,
only when a serial module is connected and is continuously
EP EA
ES
or
has been selected,
LED indicates that the control has regulated to
Control Variable
-
-
-
-
T
-
fix
Starting up
Open the device parameter menu with
the soft key
Menu .
With the cursor keys, change further to:
à
à
Control
Control Variable
Internal
is currently active.
Select other control variables (only dis-
played when present) with
and confirm with
or End ,
or quit the window with
ing any changes.
YAWE0028 / 24/06/2016
Integral XT
.
or
without mak-

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents