Thermo Scientific 7452 Operating And Maintenance Manual page 88

7452 series. cryomed controlled rate freezer
Hide thumbs Also See for 7452:
Table of Contents

Advertisement

Edit Screen
(continued)
Thermo Scientific
Figure 10-4. Ramp Rate
Step Name with Menu Format
Ramp - RAMP X.X°C/MIN UNTIL CHAMBER=Y.Y°C
RAMP X.X°C/MIN UNTIL SAMPLE=Y.Y°C
Note The slope of the temperature ramp is determined automatically,
based on the temperature entered and the chamber temperature at the
beginning of the step. Entry of a sign for the rate term is not
required/allowed.
PC Interface Software
Step Parameters with Definitions
where X.X is the ramp rate and Y.Y
is the chamber (or sample) endpoint
temperature.
Changes Chamber temp from its
current value to specified temp
(Chamber sensor trip point) at the
selected rate (°C/minute). The rate
can be set over a range of 0.1°C/
minute to 99.9°C/minute. The
temp can be set over a range of
-180°C to +50ºC(±0.1ºC). When
the Chamber reaches specified
temp, the control moves to the next
step in the profile. If Sample sensor
is selected, Chamber temp changes
at a specified rate until Sample temp
reaches specified temperature.
7007452 CryoMed Freezer
Section 10
or
10-5