Thermo Scientific 7452 Operating And Maintenance Manual page 21

7452 series. cryomed controlled rate freezer
Hide thumbs Also See for 7452:
Table of Contents

Advertisement

Section 2
Set-Up and Operation
Program User Specific
Profiles (cont.)
2-8
7007452 CryoMed Freezer
3. Press the up arrow key to view Step 0. This is the Name/Title step.
4. Press Enter to go into Edit Mode for Step 0 and the cursor will be
blinking over the leftmost character position.
To change this character, scroll through the available characters by
pressing the up or down arrow keys. Press Enter when the chosen
character is visible. The cursor will blink over the next character in the
name or title. The available characters and their relative sequence is as
follows:
! " # $ % & " ( ) * + , _ . /
0 thru 9 : ; < = > ? @
A thru Z [ ] ^ _ ' a thru z { | } → S
To delete a character, place the cursor over the chosen character and
press Silence.
Note When starting with a space 'S
a '!', while pressing the down arrow key gives you a '→'.
5. At this time, enter a unique name or title for the profile.
6. When Step 0, the name or title, is complete, move the cursor back to
the leftmost position. This is View Mode.
7. Scroll, using the down arrow key, to Step 1.
8. Press Enter to go into Edit Mode.
Step 1 contains a command with a cursor blinking just to the right of a
left pointing arrow (← __ ).
With the cursor in this position, use the up or down arrow keys to
scroll through the available commands.
p
', pressing the up arrow key gives you
p
Thermo Scientific