Keysight E4416A Installation Manual page 25

Epm-p series, peak and average power meters
Hide thumbs Also See for E4416A:
Table of Contents

Advertisement

These keys and connectors are associated with the measurement channels and
are located on the right-hand side of the front panel.
Key
Frequency
Cal Fac
Zero
Cal
Key
Keysight E4416A and E4417A Installation Guide
Function
Press this key to access the input frequency, and sensor
calibration factor menus. Use these functions to improve the
accuracy of your measurement. Refer to the user's guide for
more information.
Press this key to access the zero and calibration menus. Use
these functions to improve the accuracy of your
measurement. Refer to the user's guide for more information.
Function
The power-reference is a 1 mW (0 dBm) 50 MHz signal
POWER REF
available from a 50 ohm Type-N connector. It is used for
calibrating the sensor and meter system. If the meter is
50
MHz
configured with Option 003, the connector is fitted to the rear
panel. The green LED beside the connector is lit when the
calibrator is turned on.
The sensor input connectors (Keysight E4417A shown, the
CHANNEL
E4416A has one input). If the meter is configured with
Option 002 or Option 003, the connectors are fitted to the rear
A
panel.
B
Getting Started
25
1

Advertisement

Table of Contents
loading

This manual is also suitable for:

E4417a

Table of Contents