Operation
NOTE
Newer models of the peristaltic pump have a peg that the resting bars sit on instead
of resting on the instrument.
6
It is recommended that you switch off the water cooler.
NOTE
If the water cooler is turned off, the Peltier will automatically turn off.
Close the worksheet by clicking 'Close' from the 'File' menu but leave the
ICP Expert II software running. You may switch off the printer, monitor and any
accessories if desired.
Ensure that the powersave option on your PC is disabled (this will prevent the
shutdown of your hard disk). If this option is not disabled, you could lose data
during an unexpected shutdown.
Noxious Fumes
WARNING
Inhalation danger. The exhaust system MUST remain on if the gas supplies
are on.

Using Shutdown Mode

To shut down your instrument:
1
Rinse the spraychamber by aspirating water for a few minutes.
NOTE
When running organic samples it may be necessary to disconnect the
spraychamber and clean and dry it thoroughly between analyses.
2
Extinguish the plasma by clicking the Plasma Off icon, pressing F4 on the
keyboard or choosing Plasma Off from the Analyze menu. The peristaltic
pump stops automatically when the plasma is extinguished.
NOTE
The yellow Plasma emergency off button, located next to the mains power switch
on the front of the instrument, is designed to be used only in an emergency—for
example if the torch is melting. It is not intended to be used every time you want to
extinguish the plasma.
If used to extinguish the plasma, you will need to reset the Plasma emergency off
button to the 'on' position before the plasma can be re-ignited.
34
Agilent 700 Series ICP Optical Emission Spectrometers User's Guide