Maintenance and Troubleshooting

Axial Induction Coil

Torch
Induction coil
Torch clamp
Figure 4.
Pitch of Axial Induction Coil
Figure 5.
NOTE
Pitch is measured from the center of one coil to the next, as shown above by the
arrows between the dotted gray lines, and not the gap between adjacent coils

Cleaning

Any spills in the sample compartment should be wiped up immediately.
The user (or other authorized personnel) must perform the appropriate
decontamination procedure if hazardous material is spilled on or inside the
ICP-OES.
Agilent 700 Series ICP Optical Emission Spectrometers User's Guide
Induction coil — axial
Pitch measurement — an axial induction coil must have a pitch
(↔
) of 7.5 ±0.5 mm
41