Operation
Hot Surfaces and Noxious Fumes
WARNING
Burn danger, inhalation hazard. The plasma emits heat, ozone and fumes,
which can be hazardous. Always switch on the exhaust system before
lighting the plasma.
10 To start the ICP Expert II software, click Start > All Apps > ICP Expert II >
ICP Expert II. The Main Index window will appear.

Turning On the Instrument from Standby Mode

To turn on the instrument from Standby mode:
1
Check that all tubing on the torch, spraychamber, nebulizer and peristaltic
pump is correctly connected.
2
Close the torch compartment door, ensuring that the locking lever is fully
latched.
3
Switch on the monitor and printer (if they are off).
4
Switch on the water cooler (if it is off).
5
The Peltier will switch on automatically. If for some reason it has been turned
off, switch on the Peltier by choosing Peltier on from the Analyze menu.
NOTE
You can check if the Peltier is on by looking at the Instrument Setup window Status
page. It may take some time to reach operating temperature. It is recommended
that you wait three to five minutes before starting an analysis.
6
If you have accessories fitted, switch them on.
7
Switch on the laboratory exhaust system.
Agilent 700 Series ICP Optical Emission Spectrometers User's Guide
29