Specification Data; Specification - General; Specification - Electrical (A1 - Item 10); Specification - Electrical Connectors (A2) - GE Druck DPI 880 User Manual

Multi-function calibrator
Hide thumbs Also See for Druck DPI 880:
Table of Contents

Advertisement

Specification data

All accuracy statements are for one year.

Specification - General

Languages
English [Default]
Operating
14 to 122°F (-10 to 50°C)
temperature
Storage
-4 to 158°F (-20 to 70°C)
temperature
Humidity
0 to 90% without condensation
(Def Stan 66-31, 8.6 cat III)
Shock/Vibration
BS EN 61010:2001; Def Stan 66-31, 8.4 cat III
EMC
BS EN 61326-1:1998 + A2:2001
Safety
Electrical - BS EN 61010:2001; CE Marked
Size (L: W: H)
7.1 x 3.3 x 2.0 in
(180 x 85 x 50 mm)
Weight
15 oz (425 g)
Power supply
3 x AA alkaline batteries
mV, Volts: ≈ 60 hours
Duration
Hz, pulses: ≈ 60 hours
(Measure)
TC, mV: ≈ 70 hours
RTD, Ω: ≈ 70 hours
mA: ≈ 25 hours
mV, Volts: ≈ 50 hours
Duration
Hz, pulses: ≈ 20 hours
(Supply)
TC, mV: ≈ 70 hours
RTD, Ω: ≈ 65 hours
mA: ≈ 10 hours (24 V Source at 12 mA)

Specification - Temperature ranges (RTD)

RTD type
Standard
Pt50 (385)
IEC 751
Pt100 (385)
IEC 751
Pt200 (385)
IEC 751
Pt500 (385)
IEC 751
Pt1000 (385)
IEC 751
D 100 (392)
JIS 1604-1989
Ni 100
DIN 43760
Ni 120
MINCO 7-120
*Temperature coefficient:
14 to 50°F, 86 to 122°F = 0.0028% FS / °F
(-10 to 10°C, 30 to 50°C = 0.005% FS / °C)

Specification - Resistance ranges (Ohms/RTD)

Range (Ω)
Excitation (mA)
0 to 400 0.1 to 0.5
0 to 400 0.50 to 3.0
400 to 1500 0.10 to 0.8
1500 to 3200 0.05 to 0.4
3200 to 4000 0.05 to 0.3
*Temperature coefficient:
14 to 50°F, 86 to 122°F = 0.0028% FS / °F
(-10 to 10°C, 30 to 50°C = 0.005% FS / °C)
K405 Issue 1
Range °F
Range °C
-328 to 1562
-200 to 850
-328 to 1562
-200 to 850
-328 to 1562
-200 to 850
-328 to 1562
-200 to 850
-328 to 752
-200 to 400
-328 to 1202
-200 to 650
-76 to 482
-60 to 250
-112 to 500
-80 to 260
Accuracy (Ω)*
Measure
Supply
-
0.15
0.10
0.10
0.50
0.50
1.00
1.00
1.30
1.30

Specification - Electrical (A1 - Item 10)

Range (Measure):
Accuracy: Measure mA
Accuracy: Measure mV
Accuracy: Measure V
Range (Supply):
Accuracy (Supply):
mA, mV, V
Temperature coefficient
(Measure or supply)
14 to 50°F, 86 to 122°F
(-10 to 10°C, 30 to 50°C)
Connectors (A1 - Item 10)
*
Refer to "Specification - Resistance ranges (Ohms/RTD)"

Specification - Electrical connectors (A2)

Range (Measure)
Accuracy: Measure mA
Accuracy: Measure V
Temperature coefficient
14 to 50°F, 86 to 122°F
(-10 to 10°C, 30 to 50°C)
Switch detection
Loop power output
®
HART
resistor
Connectors (A2)
Accuracy °F *
0.90
0.45
1.08
0.72
0.36
0.45
0.36
0.36
0 to ±55 mA
0 to ±120 mV
0 to 4000Ω*
0 to ±30 V
0.02% of reading + 3 counts
0.02% of reading + 2 counts
0.03% of reading + 2 counts
0 to 24 mA
0 to 120 mV
0 to 4000Ω*
0 to 12 V
0.02% of reading + 2 counts
0.0017% FS / °F
(0.003% FS / °C)
Four 0.16 in (4 mm) sockets
One TC mini-connector socket
0 to ±55 mA
0 to ±30 V
0.02% of reading + 3 counts
0.03% of reading + 2 counts
0.0017% FS / °F
(0.003% FS / °C)
Open and closed. 2 mA current.
24 V ± 10% (Maximum: 35 mA)
250 Ω
Three 0.16 in (4 mm) sockets
Accuracy °C *
0.50
0.25
0.60
0.40
0.20
0.25
0.20
0.20