The models in NISE 3640M series are intended for use in the electromagnetic
environment specified below. The customer or the user of the models
in NISE 3640M series should assure that it is used in such an environment.
Immunity Test
IEC 60601 Test Level
Conducted RF
3 Vrms
IEC 61000-4-6
150 kHz to 80 MHz
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to 2,5 GHz
Electrostatic discharge (ESD)
±6 kV contact
IEC 61000-4-2
±8 kV air
Electrical fast
±2 kV for power
transient/burst
supply lines
±1 kV for input/output
IEC 61000-4-4
lines
Surge
±1 kV line(s) to line(s)
IEC 61000-4-5
±2 kV line(s) to earth
Copyright © 2013 NEXCOM International Co., Ltd. All Rights Reserved.
Compliance Level
interruptions and voltage
3 Vrms
variations on power supply
input lines
3 V/m
IEC 61000-4-11
±2 kV ±4 kV ±6 kV contact
±2 kV ±4 kV ±8 kV kV air
±2 kV for power
supply lines
±1 kV for input/output
lines
±1 kV line(s) to line(s)
Power frequency
±2 kV line(s) to earth
(50/60 Hz) magnetic field
NOTE: UT is the a.c. mains voltage prior to application of the test level.
xiii
Immunity Test
IEC 60601 Test Level
0 % UT (>95 % dip in UT)
for 0,5 cycle
40 % UT (60 % dip in UT)
for 5 cycles
70 % UT (30 % dip in UT)
for 25 cycles
0 % UT (>95 % dip in UT)
for 5 sec
3 A/m
Compliance Level
0 % UT (>95 % dip in UT)
for 0,5 cycle
40 % UT (60 % dip in UT)
for 5 cycles
70 % UT (30 % dip in UT)
for 25 cycles
0 % UT (>95 % dip in UT)
for 5 sec
3 A/m