Thermo Scientific Orion Dual Star User Manual page 67

Ph/ise dual channel benchtop meter
Hide thumbs Also See for Orion Dual Star:
Table of Contents

Advertisement

Thermo Scientific
Note: The printer must have a 9600 baud rate to be connected to the autosampler. The Orion
Star series printer, Catalog Number 1010006, is recommended for use with the Orion Dual Star
meter and autosampler.
4. Prepare and install the electrodes. Prepare the electrodes according to the electrode user
guides. Connect the electrodes to the meter using the extension cables and cable
management system provided with the autosampler. Insert the electrodes into the electrode
holder on the autosampler. If a stirrer probe will be used, connect the stirrer probe to the
stirrer jack on the back of the meter and insert the stirrer probe into the electrode holder on
the autosampler.
Note: The minimum cable length required to connect electrodes to the autosampler is 6.25 feet
(1.9 meters). Extension cables or electrodes with an appropriate cable length are needed.
5. Connect the meter, autosampler and printer (if applicable) to a power supply and turn on the
power. The default Orion Dual Star meter baud rate is 9600 and the default autosampler
baud rate for the COM 1 port is 1200, so the meter baud rate needs to be set to 1200 or the
autosampler dip switches need to be adjusted to 9600 for the COM 1 port.
Autosampler Operation
Orion Dual Star pH/ISE Benchtop Meter User Manual
Section 8
|
67
|