When the meter displays Keypad Test Passed, press the f2 (next) key. The meter will
display Self Test Passed. Press the f1 (escape) key to return to the measurement mode.
User Reset – Press the ▲ / ▼ keys to highlight User Reset and press the f2 (accept) key.
•
If the user reset is performed, all of the current meter settings will be deleted and reset to
the default settings. The data log, calibration log and methods will not be deleted with the
user reset. Press the f2 (yes) key to perform the reset or press the f3 (no) key to abort the
reset. The meter will enter the EZ Startup menu and then proceed to the measurement
mode.
Factory Config – Press the ▲ / ▼ keys to highlight Factory Config and press the
•
f2 (accept) key. If the factory configuration is performed, all of the current meter settings will
be deleted and reset to the default settings and the data log, calibration log and methods
will be deleted. Press the f2 (yes) key to perform the reset or press the f3 (no) key to abort
the reset. Use the alphanumeric keypad to enter the PIN number and press the f2 (accept)
key to perform the reset. The default PIN number is 111111111. Press the f1 (measure)
key, the meter will enter the EZ Startup menu and then proceed to the measurement mode.
About Meter – Press the ▲ / ▼ keys to highlight About Meter and press the f2 (accept)
•
key. The meter serial number and software version will be displayed. Press the f1 (back)
key to return to the troubleshooting settings and press the f1 (back) key to return to the
instrument parameter settings.
Beeper Volume
Press the ▲ / ▼ keys to highlight Beeper Volume and press the f2 (select) key. Press the
▲ / ▼ keys to highlight Off, Low or High and press the f2 (accept) key.
Language
Press the ▲ / ▼ keys to highlight Language and press the f2 (select) key. Press the ▲ / ▼
keys to highlight English, Español, Deutsch, Français or Italiano and press the f2 (accept) key.