Section 3
Meter Setup
|
26
Orion Dual Star pH/ISE Benchtop Meter User Manual
|
Measurement or Timed Reading as the read type that will be used to display the measurements
for the selected channel and press the f2 (select) key.
Auto-Read – The meter will display the measurement as it stabilizes and lock and hold the
measurement when it is stable. Once a measurement is locked, press the measure key to
take a new measurement. Once the measurement is locked, the meter will automatically
export the measurement to the data log, if the data log is enabled in the setup menu, and
to a printer or computer, if a printer or computer is connected to the meter and enabled in
the setup menu.
On Ready – The meter will show stabilizing on the display until the measurement is stable.
When the measurement is stable, the meter will show ready on the display. The display will
automatically update if the measurement changes and the measurement will not lock.
Once ready is shown on the display, the meter will automatically export the measurement
to the data log, if the data log is enabled in the setup menu, and to a printer or computer, if
a printer or computer is connected to the meter and enabled in the setup menu.
At Time Intervals – The meter will read and display a measurement at the set time interval
that is programmed by the operator. At the set time interval, the meter will automatically
export the measurement to the data log, if the data log is enabled in the setup menu, and
to a printer or computer, if a printer or computer is connected to the meter and enabled in
the setup menu. If At Time Intervals is selected, enter the time interval value in a hours :
minutes : seconds format using the numeric keypad and press the f2 (accept) key.
Continuous – The meter will continuously measure and update the display. No indicator will
be shown on the display. This read type is useful when performing an experiment that
requires continuous measurements to be taken, regardless of the measurement stability.
Press the log/print key to export the measurement to the data log, if the data log is enabled
in the setup menu, and to a printer or computer, if a printer or computer is connected to the
meter and enabled in the setup menu.
Value Change In Measurement – The meter will display the measurement when the
reading reaches or exceeds the set high or low value that is programmed by the operator.
When the reading reaches or exceeds the set high or low value, the meter will
automatically export the measurement to the data log, if the data log is enabled in the
setup menu, and to a printer or computer, if a printer or computer is connected to the meter
and enabled in the setup menu. If Value Change In Measurement is selected, press the
▲ / ▼ keys to highlight Min Value, enter the minimum measurement value using the
numeric keypad and the decimal key, press the ▲ / ▼ keys to highlight Max Value, enter
the maximum measurement value using the numeric keypad and the decimal key and
press the f3 (next) key.
Timed Reading – The meter will display a measurement after a set time delay and lock and
hold the measurement after the time delay is reached. To start another time delay cycle,
press the measure key. Once the measurement is locked, the meter will automatically
export the measurement to the data log, if the data log is enabled in the setup menu, and