Vega VEGAPULS 63 Operating Instructions Manual page 3

Radar sensor for continuous level measurement of liquids, 4...20 ma/hart two-wire
Hide thumbs Also See for VEGAPULS 63:
Table of Contents

Advertisement

Parameterization example
Further steps
VEGAPULS 63 • 4 ... 20 mA/HART two-wire
4. Carry out the adjustment in the menu items "Min. adjustment" and
"Max. adjustment".
The radar sensor measures the distance from the sensor to the
product surface. For indication of the real level, an allocation of the
measured distance to the percentage height must be carried out.
Fig. 3: Parameter adjustment example min./max. adjustment
1
Min. level = max. meas. distance
2
Max. level = min. meas. distance
For this adjustment, the distance is entered when the vessel is full
and nearly empty. If these values are not known, an adjustment with
other distances, for example, 10 % and 90 % is also possible. Starting
point for these distance specifications is always the seal surface of
the thread or flange.
1. In the menu "Additional settings", menu item "Damping" you can
adjust the requested damping of the output signal.
2. Select the output characteristics in the menu item "Current out-
put".
The quick start is then finished. For further information see chapter
"Parameter adjustment".
Quick start
2
100%
1
0%
3

Advertisement

Table of Contents
loading

Table of Contents