Vega VEGAPULS 63 Operating Instructions Manual page 19

Radar sensor for continuous level measurement of liquids, 4...20 ma/hart two-wire
Hide thumbs Also See for VEGAPULS 63:
Table of Contents

Advertisement

VEGAPULS 63 • 4 ... 20 mA/HART two-wire
100%
0%
Fig. 14: Configuration surge pipe
1
Radar sensor
2
Marking of the polarisation direction
3 Thread or flange on the instrument
4
Vent hole
5
Holes
6
Weld joint
7 Welding neck flange
8
Ball valve with complete opening
9
Fastening of the surge pipe
Instructions for orientation:
Note marking of the polarisation plane on the sensor
With threaded fitting, the marking is on the hexagon, with flange
connection between the two flange holes
All holes in the surge pipe must be in one plane with this marking
Instructions for the measurement:
The 100 % point must be below the upper vent hole
The 0 % point is the end of the surge pipe
For the parameter adjustment, select "Application standpipe" and
enter the tube diameter to compensate errors due to running time
shift
A false signal suppression with integrated sensor is recommended
but not mandatory
The measurement through a ball valve with complete run is pos-
sible