Circuit Breaker
2.4

Moving the Unit

Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
The system provides an automatic circuit breaker, for
emergency situations, such as power surges, that could
damage the unit.
If the circuit breaker trips:
1. Unplug the unit.
2. Press the white button, on the left side of the unit.
3. Plug the unit back in.
Suction cups, at the bottom of the unit, keep it anchored to
the work surface. Keeping the unit stationary is a safety
feature.
To move the unit to a new location:
Caution: The unit can weigh up to 145 lb. (65.8 kg). Use
precaution when moving to avoid any injury.
1. Check that the new site meets the criteria in Section 2.2
before moving the unit.
2. Position a flat object, such as a tongue depressor, near a
suction cup at the bottom of the unit.
3. Lift up an edge of the cup, and insert the flat object far
enough to break the vacuum seal.
4. When all four cups are disengaged, lift the unit from the
work surface.
Caution: When the unit is in its new location, ensure that
the suction cups adhere correctly to the work surface.
MARATHON Series Operation Manual
10