5.3

Cover Interlock Bypass

5.4

Fuses not replaceable by operator

Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
The cover will remain locked, if power fails. If you need to
remove samples from the unit, before power is restored, use
the cover interlock bypass, after the rotor has come to a
stop.
To bypass the cover interlock:
1. Unplug the centrifuge.
2. Locate the hidden plastic plug, underneath the front
ledge of the cabinet.
3. Use a screwdriver to pry out and remove the plug.
4. Pull the attached cord, to release the cover interlock.
5. Replace the plug in the hole.
Do not perform this operation routinely. The centrifuge's
cover interlock provides operator safety. It allows the
cover to be opened promptly, whenever rotation has
stopped.
The following fuses are located internally within the
centrifuge on the power PC board. They should only be
replaced by qualified service personnel.
F1
10A
Type F Fast Blow
F2
1A
F3
1A
MARATHON Series Operation Manual
39
250V
Fast Blow
250V
Fast Blow
250V