Decontamination Table - Fisher Scientific MARATHON 3000 Operation Manuals

General purpose centrifuge
Table of Contents

Advertisement

4.5

Decontamination Table

Compatible Processes For Decontamination
Sterilization Methods
Mechanical
Autoclave*
Ethylene Oxide Gas
Dry Heat (2Hrs. @ 160°C)
Chemical
Ethanol
40% Formalin
Methanol
2-Propanol
.5% Sodium Hypochlorite**
3% Hydrogen Peroxide
100% Hydrogen Peroxide
5% Phenol Solution
*For Oxidizing Acids, see "Oxidizing Agents, strong".
PA - POLYALLOMER
PC - POLYCARBONATE
PE - POLYETHYLENE
PP - POLYPROPYLENE
PU - POLYURETHANE
NL - MODIFIED PHENYLENE OXIDE (NORYL)
DN - ACETAL HOMOPOLYMER (DELRIN)
CN - ACETAL COPOLYMER (CELCON)
NN - NYLON
PS - POLYSTYRENE
Warning:
This chart describes the material compatibility of various sterilization methods. It does
not specify the adequacy of sterilization. Refer to section 4.4 - Chemical Resistance
Table, for material compatibility during centrifugation.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Plastic
PA PC PE PP PU NL DN CN NN PS TI SS AL MB MG RR BN VN PF PT
S
M
U
S
M
U
S
S
S
S
S
S
S
S
U
U
U
U
U
U
U
S
S
S
S
U
S
S
S
S
S
S
U
S
S
S
M
S
S
M
S
S
S
S
S
S
M
S
S
S
S
S
S
U
S
U
S
S
S
S
S
S
M
S
S
S
S
S
U
U
M
U
U
S
U
U
M M
TI - TITANIUM
SS - STAINLESS STEEL
AL - ALUMINUM
MB - MANGANESE BRONZE
MG - MAGNESIUM
RR - RUBBER
BN - BUNA-N
VN - VITON
PF - PHENOLIC FIBER
PT - PAINTED SURFACES
MARATHON Series Operation Manual
34
Metal
S
S
U
S
S
S
S
S
S
S
S
S
S
S
U
U
U
S
S
U
S
S
U
M
S
S
S
S
S
S
U
S
S
S
S
S
U
M
S
S
S
S
S
U
S
S
S
S
S
U
U
S
S
M
U
U
S
U
S
S
S
S
S
U
U
S
S
S
S
S
U
M M M M M M M
*Autoclaving
121°C for 20 min.
@ 2 ATM (15 PSIG)
**1 to 10 Dilution of
Household Bleach
S=
M=MARGINAL
U=UNSATISFACTORY
Other
S
S
S
M
S
M
S
U
U
S
S
S
S
U
U
U
U
U
S
S
S
S
S
S
S
S
U
S
S
S
S
S
S
U
S
S
M
S
S
S
S
S
U
S
U
S
S
M
U
S
S
S
S
M
S
U
U
S
S
U
U
S
S
U
SATISFACTORY

Advertisement

Table of Contents
loading

Table of Contents