5.
Operation

5.4.6 Trimming and sectioning functions

Switching between trimming
and sectioning
Press this key to alternate between the
two functions.
illuminated:
Trimming mode is active
off:
Sectioning mode is active.
Selecting trimming thickness
• Activate trimming mode
• Select trimming thickness (= sectioning
thickness at which trimming is carried out)
by pressing
or
The selected thickness is displayed in the cor-
responding LED
Motorized trimming
• Select 'Continuous stroke' in functional area
A.
• Activate trimming by pressing RUN/STOP
and RUN/ENABLE (located in functional
area C) simultaneously or by depressing the
footswitch.
• To stop trimming press RUN/STOP or RUN/
ENABLE or depress the footswitch again.
Trimming thickness can be changed while trim-
ming is in motion by pressing
Leica CM 3050 S – Cryostat
Functional area B - all models
.
:
or
.

5.4.7 Setting the section thickness

• Activate sectioning mode (
C T . . . . ° C
• Press
or
to select section thickness.
Pressing one of these keys once makes the
display in control panel 1 switch to section
thickness indication (thickness value not yet
being altered). Pressing
starts to change the actual section thick-
ness setting.
Selectable section thickness settings:
0 -
2 μm
in 0.5-μm
2 -
10 μm
in 1-μm
10 -
20 μm
in 2-μm
20 -
60 μm
in 5-μm
60 - 100 μm
in 10-μm
100 - 300 μm
in 50-μm
off).
O T . . . . ° C
μ 0 1 0 . 0
or
further
steps,
steps,
steps,
steps,
steps,
steps.